Table 4
Uncertainties for neon of the inter-comparisons CCT-K2 and CCT-K2.x with and without the contribution of the isotopic factor.
Comparisons | # | Cell fabricated (measured) | Sealing date | Cell | UKC,original a | Ulab meanb | UCORR meanc | Better byi | |||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
#1 | #2 | Uiso onlyd | U (Type B) | UTOTlabe | UNET without Uisof | Uiso,fg | UTRUEh | ||||||
μK | |||||||||||||
K2, Star | 1 | BNM-INM | 1985 | BCMH2O | 1080j | 2800j | – | 1100j | 1100j | – | – | – | |
K2, Int84, Star | 2 | CNR-IMGC | 1979 | 3Ne | 280 | 280 | 110 | 130 | 210 | 170 | 22 | 171 | 39% |
K2 | 3 | KRISS | – | Open | 400 | 400 | 300n | 340 | 360 | 100 | 22 | 102 | 74% |
K2 | 4 | NIST | – | Open | 320 | 320 | 150 | 180 | 265 | 170 | 22 | 171 | 46% |
K2, Star, INRIM | 5 | NPL | 1993 | Ne2 | 440 | 380 | 300 | 190 | 200 | 80 | 6 | 80 | 79% |
K2, K2.1 | 6 | NRC | 1985 | F15 | 440 | 440 | 320n | 360 | 400 | 120 | 1 | 120 | 73% |
K2, Star, INRIM | 7 | PTB | 1995 | Ne-7 | 400 | 320 | 340 | 360 | 80 | 72 | 108 | 73% | |
K2.1 | 8 | VNIIFTRI | – | Open | 670 | 400n | – | 420 | 270 | ||||
K2.3, Star | 9 | INRIM (VSL) | 1999 | 12Ne | 620 | 620 | 314 | – | 392 | 306 | 48 | 310 | 50% |
K2.4, Star | 10 | INM | 2002 | Ne02/1 | 760 | 760 | 480 | 600 | 600 | 280 | 72 | 289 | 62% |
K2.4 | 11 | INRIM (INTiBS) | 2000 | E3Ne | 640 | 640 | 320n | – | 400 | 320 | 48 | 324 | 49% |
K2.3, K2.4 | 12 | NRC | 2004 | Cu-M-1 | 640 | 640 | 320 | – | 400 | 320 | 94 | 334 | 48% |
K2.3, K2.4, Star | 13 | NRC | 1985 | F17 | 640 | 640 | 320 | – | – | 320 | 2 | 320 | 50% |
K2.5, INRIM | 14 | NMIJ-AIST | 2006 | Ne-5 | 600 | 600 | 280o | – | 290 | 320 | 8 | 320 | 50% |
K2.5, (K2.2), INRIM | 15 | INRIM | 2002 | Ec2Ne | 680 | 680p | 8q | – | 110 | – | 16 | – | – |
s.d.k | 15%m | ||||||||||||
Meanl | 60%m |
Original uncertainties of the ΔTs (ΔTor in Table 2).
Original laboratory uncertainty budgets from [19]: mean of #1 and #2 sets.
Uncertainties (a) cleaned up from the isotopic uncertainties (“Uiso only” under (b)): mean on #1 and #2 sets.
Uncertainty of the isotopic corrections taken from [15].
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.