Open Access

Table 4

Uncertainties for neon of the inter-comparisons CCT-K2 and CCT-K2.x with and without the contribution of the isotopic factor.

Comparisons # Cell fabricated (measured) Sealing date Cell UKC,originala Ulab meanb UCORR meanc Better byi



          #1 #2 Uiso onlyd U (Type B) UTOTlabe UNET without Uisof Uiso,fg UTRUEh  
          μK  
K2, Star 1 BNM-INM 1985 BCMH2O 1080j 2800j 1100j 1100j  
K2, Int84, Star 2 CNR-IMGC 1979 3Ne 280 280 110 130 210 170 22 171 39%
K2 3 KRISS Open 400 400 300n 340 360 100 22 102 74%
K2 4 NIST Open 320 320 150 180 265 170 22 171 46%
K2, Star, INRIM 5 NPL 1993 Ne2 440 380 300 190 200 80 6 80 79%
K2, K2.1 6 NRC 1985 F15 440 440 320n 360 400 120 1 120 73%
K2, Star, INRIM 7 PTB 1995 Ne-7   400 320 340 360 80 72 108 73%
K2.1 8 VNIIFTRI Open 670   400n 420 270      
K2.3, Star 9 INRIM (VSL) 1999 12Ne 620 620 314 392 306 48 310 50%
K2.4, Star 10 INM 2002 Ne02/1 760 760 480 600 600 280 72 289 62%
K2.4 11 INRIM (INTiBS) 2000 E3Ne 640 640 320n 400 320 48 324 49%
K2.3, K2.4 12 NRC 2004 Cu-M-1 640 640 320 400 320 94 334 48%
K2.3, K2.4, Star 13 NRC 1985 F17 640 640 320 320 2 320 50%
K2.5, INRIM 14 NMIJ-AIST 2006 Ne-5 600 600 280o 290 320 8 320 50%
K2.5, (K2.2), INRIM 15 INRIM 2002 Ec2Ne 680 680p 8q 110 16
s.d.k                       15%m
Meanl                       60%m
a

Original uncertainties of the ΔTs (ΔTor in Table 2).

b

Original laboratory uncertainty budgets from [19]: mean of #1 and #2 sets.

c

Uncertainties (a) cleaned up from the isotopic uncertainties (“Uiso only” under (b)): mean on #1 and #2 sets.

d

Item for the isotopic effect of the estimated laboratory uncertainty.

e

Total laboratory uncertainty budget.

f

Uncertainty budget corrected for the isotopic effect contribution.

g

Uncertainty of the isotopic corrections taken from [15].

h

New total uncertainty budget taking also (f) into account.

i

Lowering of (h) with respect to (e).

j

Data not elaborated.

k

Standard deviation of the column data.

l

Mean of the column data.

m

Improvement of (h) on (e) in percent.

n

Also includes chemical impurities.

o

Uncorrected, used in the uncertainty budget.

p

For the direct comparison (INRIM − NRC) 300 μK.

q

Uncertainty of the corrected value.