Open Access

Table 2

Data used in this study from comparisons involving the Neon triple point, T90tp = 25.5661 K. Basic data.

# Data from comparisona Sample fabricated (owned by) Year (S = sealed) Cell #b Gas (assay #) c Purity, nom. (measured) Set #1, CCT-K2.x comparison resultsd, Set#2    

              ΔToriginal/mK U/mK ΔToriginal/mK U/mK
1 K2, Star BNM-INM 1985 (S) BCMH2O AL 4N −0.02 1.08 −1.88 2.8
2 K2, Int84, Star CNR-IMGC 1979 (S) 3Ne M-b (#2) 4N5 (>4N5) 0.11 0.28 0.11 0.28
3 K2 KRISS <1997 Openg M 4N4 0.01 0.4 −0.15 0.4
4 K2 NIST 1979 Open M (#8) >5N5 –0.13 0.32 0.04 0.32
5 K2, Star, INRIM NPL 1993 (S) Ne2 AP (#18) >5N –0.10 0.44 −0.19 0.38
6 K2, K2.1 NRC 1985 (S) F15 AP (#17) 5N –0.06 0.44 −0.12 0.44
7 K2, Star, INRIM PTB 1995 (S) Ne-7 L (#10) 5N 0.26 0.4
8 K2.1 VNIIFTRI Open 0.28 0.67 0.22 0.67
9 K2.3, Star INRIM (VSL) 1999 (S) 12Ne MG (#11) 5N −0.645 0.66 −0.71 0.66
10 K2.4, Star INM 2002 (S) Ne02/1 AL (#14) 5N (4N2) −0.625 0.78 −0.685 0.78
11 K2.4 INRIM (INTiBS) 2000 (S) E3Ne MG (#11) 5N −0.245 0.64 −0.305 0.64
12 K2.3, K2.4, K2.5 NRC 2004 (S) Cu-M-1 P (#9) 5N −0.225 0.20 −0.285 0.20
13 K2.3, K2.4, K2.5, Star NRC 1985 (S) F17 AP (#17) 5N −0.133 0.20 −0.173 0.20
14 K2.5, INRIM NMIJ-AIST 2006 (S) Ne-5 AW (#7) 5N (4N2) −0.54f 0.60 −0.60f 0.61
15 K2.5, (K2.2), INRIM INRIM 2002 (S) Ec2Ne MG (#11) 5N −0.69 0.68 −0.75 0.68
16 (K2.2), Star INRIM (NIM) 2000 (S) E2Ne MG (#11) 5N K2.2 not yet completed      
17 Int84, Star, INRIM CNR-IMGC 1977 (S) 1Ne M-a (#3) 4N5        
18 Star INRIM (PTB) 1999 (S) E1Ne MG (#11) 5N        
19 Star INRIM (DSIR) 1986 (S) 11Ne S (#13) 4N        
20 Star INRIM (INTiBS) 2002 (S) 7Ne S (#13) 4N        
21 INRIM NMIJ 2005 (S) Ne-2 IB (#5) 5N        
22 INRIM INRIM 1999 (S) 15Ne MG (#11) 5N        
23 INRIM INRIM 2000 (S) E4Ne MG (#11) 5N        
24 INRIM INRIM 2001 (S) Ec1Ne MG (#11) 5N        
25 Int84, Star VNIIFTRI 1997 (S) MC-897 own 5N        
26 Int84 NRC 1979 (S) Cell 12 M 4N5        
27 Int84 INM 1982 (S) BCM4 AL 4N        
28 Int84 NRLM 1978 (S) 1Ne J 4N        
29 Int84 NRLM 1978 (S) 2Ne J 4N        
a

K2 = CCT-K2 (NRC Pilot, 1997–99) [19]; K2.1–K2.5 = bi- and tri-lateral CCT-K2.x (NRC Pilot, 2003–2015) [2426]; Star = Comparison (PTB Co-ordinator, 2007–2012) [12]; Int84 = International Intercomparison (IMGC-CNR Pilot, 1978–84) [34]; INRIM = Comparison at INRIM (INRIM Co-ordinator, 2008–2009) [37,38].

b

Samples measured.

c

Assays reported in [15].

d

Original differences with respect to the KCRV from [19]. Set#: indicates thermometers set.

f

Mean of results.

g

Measurements repeated in 2013 with the isotopic composition known and corrected for it; the same thermometer has measured samples from two other bottles.