Open Access

Table 2

Sources of uncertainties associated with their values.

Sources of uncertainties Standard uncertainty, ui Probability distribution Sensitivity coefficient, ci Uncertainty (N/mm2)
Calibration, uF1 590.4 [N] Rectangular 1.61 × 10−5 [1/mm2] 9.49 × 10−3
Load application rate, uF2 2778.9 [N] Rectangular 1.61 × 10−5 [1/mm2] 4.46 × 10−2
Preservation condition effect, uF3 2084.2 [N] Rectangular 1.61 × 10−5 [1/mm2] 3.35 × 10−2
Effect of planarity, uF4 2084.2 [N] Rectangular 1.61 × 10−5 [1/mm2] 3.35 × 10−2
Angle of surface plane, uF5 138.9 [N] Rectangular 1.61 × 10−5 [1/mm2] 2.23 × 10−3
Sample centering, uF6 694.7 [N] Rectangular 1.61 × 10−5 [1/mm2] 1.11 × 10−2
Reading of applied force, uF7 105.1 [N] Normal 1.61 × 10−5 [1/mm2] 1.69 × 10−3
Reading of sample area, uA 73.4 [mm2] Normal 3.59 × 10−5 [N/mm4] 2.64 × 10−3
Combined uncertainty uc(y), k = 1 1.39 × 10−1
Expanded uncertainty U, k = 2 2.78 × 10−1

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.