Research Article
Evaluation of InGaAs/InP photodiode for high-speed operation at 4 K
1
University of South-Eastern Norway,
Borre, Norway
2
Justervesenet,
Kjeller, Norway
* Corresponding author: eivind.bardalen@usn.no
Received:
11
December
2017
Accepted:
14
October
2018
An optically controlled high-speed current source located at 4 K is likely to improve the performance of pulse-driven Josephson junction arrays. A custom photodiode module with an Albis PDCS24L InGaAs/InP PIN photodiode is investigated in order to determine the suitability at 4 K. The DC and frequency response were tested at room temperature and at temperatures down to 4 K. For continuous wave optical input, photocurrents above 15 mA were produced at both room temperature and 4 K. I–V measurements show that the threshold voltage increased from 0.5 V at room temperature to 0.8 V at 4 K. The transmission coefficient S21 of the optoelectronic system, including cables and modulated laser source, was measured using a vector network analyzer. Up to 14 GHz, the results showed that the frequency response at 4 K was not degraded compared to room temperature. At room temperature, reverse bias voltages of up to 3 V was required for the highest bandwidth, while at 4 K, the photodiode was operated at nearly full speed even at 0 V reverse bias.
Key words: optoelectronics / cryogenics / voltage standards
© E. Bardalen et al., published by EDP Sciences, 2018
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.