Open Access

Table 2

Results from calibration of the laboratory setup for measuring contact angle. This result was also communicated in a report of the MFMET project [33].

Angle of standard (°) Measured angle (°) Error (°)
30 30.55 0.55
30 29.35 –0.65
30 29.77 –0.23
60 59.84 –0.16
60 59.70 –0.30
60 59.32 –0.68
120 119.932 –0.068
120 120.20 0.20
120 120.13 0.13
Average of errors E¯ (°) –0.13
Standard deviation of errors σE (°) 0.40