Research Article
Quality approach towards ISO/IEC 17025 accreditation at the center of analysis and characterization of Marrakesh Cadi Ayyad University: metrology stakes
1
Laboratory of Applied Chemistry and Environment, Faculty of Sciences and Technologies, Hassan I University, Settat, Morocco
2
Faculty of Sciences and Technologies, Hassan I University, Settat, Morocco
3
Laboratory of Water, Biodiversity and Climate Change, Faculty of Sciences Semlalia, Cadi Ayyad University, Marrakesh, Morocco
4
Laboratory of Materials, Energy and Environment, Faculty of Sciences Semlalia, Cadi Ayyad University, Marrakesh, Morocco
5
CIRAD, DGDRS, 34398 Montpellier, France
6
Montpellier University, Montpellier, France
7
CIRAD, UPR BioWooEB, 34398 Montpellier, France
8
BioWooEB, Univ Montpellier, CIRAD, Montpellier, France
9
QMP Management, Marrakesh, Morocco
10
Polytech Angers − LARIS, Angers University, France
11
Cadi Ayyad University, Marrakesh, Morocco
* Corresponding author: marie-france.thevenon@cirad.fr
Received:
21
December
2023
Accepted:
15
September
2024
In the context of the European Tempus project “Quality in Higher Agricultural Education in the Mediterranean (QESAMED)”, Analysis and Characterization Center of Cadi Ayyad University has committed to implement a quality approach according to the ISO/IEC 17025 standard. The first objective was to accredit the testing carried out in the microbiology laboratory in response to strong demand from the food industry in the Marrakesh region. The process began with an initial assessment of the center activities to identify the main gaps from ISO/IEC 17025. The range of laboratory and standard testing to be accredited were determined and an appropriate action plan was established. After that, a staff training was programmed to improve their skills in relation to this standard and metrology concepts. Implementation of the metrology function is one of the key steps for the deployment of the continuous improvement process. This function guarantees the traceability of measurements and the reliability of microbiological testing results. Several actions have been carried out, including: (i) identification of critical quantities and associated metrological requirements, (ii) checking of the metrological consistency of equipment through the calculation of the capability coefficient. The management of equipment requires several steps, from receipt of the equipment to its decommissioning or reform: (i) identification of the equipment, (ii) creation of equipment files, (iii) performance of calibration, verification, and maintenance operations. Ultimately, the definition of a strategy ensuring metrological traceability will optimize management costs by taking into account the structural constraints linked to the organization of metrology on a national scale.
Key words: ISO/IEC 17025 / equipment / metrological requirements / capability / traceability
© M. Outaki et al., Published by EDP Sciences, 2024
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1 Introduction
Cadi Ayyad University (CAU) was created in 1978 in Marrakesh (Morocco). It currently includes 15 establishments in 4 cities, namely: Marrakesh, Kalaa des Sraghna, Essaouira and Safi, and brings together 1596 scientific staff (professors), 801 administrative staff, and more than 95186 students [1]. As part of the development of its research strategy, the CAU created, among others, the Center of Analysis and Characterization (CAC) attached to “Innovation City” (the CAU technological platform) whose goals are to optimize and pool research equipment, pool resources, and boost university-business partnerships [1].
The CAC activities started in 2011 within the CAU Faculty of Sciences, and was relocated in 2013 into a dedicated new building covering an area of 1780 m2 on two floors and housing 13 laboratories, with dedicated measurement and testing equipment. The purpose of the center is to provide testing services in the fields of nanoscience, molecular biology, and agri-food. The CAC performs thousands of testing every year.
The CAC agri-food platform includes equipment used for the physicochemical characterization of foods through the techniques of chromatography, spectrophotometry, polarimetry, etc. The creation of a laboratory specializing in microbiological analyses meets the needs of agri-food companies on a regional, as well as national levels, which, further to the development of hygiene and food safety standards and the increasingly regulatory and consumer requirements, need to carry out quality controls to prove the wholesomeness of their products.
In today's world, quality management is the only route possible for African countries wishing to meet the challenge of globalization [2]. Implementing a quality approach in research laboratories, and especially at the level of a university platform, is currently a major stake to guarantee the results provided and ensure the sustainability of research activities [3,4]. In Africa, there is a real need to develop and maintain skills in metrology and quality management. This point is considered by some authors to be a great challenge [2]. Other authors say that implementing a quality system compatible with ISO/IEC 17025 in a university is entirely feasible and adds significant value to the university [5,6], but it remains a difficult task [7].
Under the impetus of the partnership developed between CAU and the consortium of the European project “Quality in Higher Agricultural Education in the Mediterranean (QESAMED1)”, the CAC has integrated the implementation of this approach into its objectives, with the aim of first accrediting the CAC's microbiology laboratory testing according to the requirements of the standard ISO/IEC 17025 [8]. The latter implies the need to set up a metrology function to ensure the reliability and traceability of testing results. The control of the various measurement and testing equipment becomes mandatory and requires performing several types of operations as calibration, checks, preventive and curative maintenance, etc. Metrology has a preventive role that guarantees high standards for analytical results and gives trust in the conclusions reached [9]. If it is not implemented in certain laboratories it is because that it is a demanding task requiring significant investment [10]. This investment covers the price of buying the necessary equipment, training the staff, and managing the time required for implementing the metrology function. Additionally, it includes ensuring its sustainability through periodic controls [11].
The purpose of this paper is to describe the approach followed to set up the metrology function at the CAC, and more particularly in the microbiological testing laboratory, which will serve as an example for the other laboratories of the platform. A focus is given on the main difficulties encountered and on the solutions that could help overcome them.
2 Methods
The quality approach at the CAC was initiated after the selection of the center as a pilot process in the QESAMED project. The latter aims to strengthen the capacities of research and higher education establishments in Morocco, Algeria, Tunisia, and Lebanon by improving their agronomic training courses. The project's vision essentially includes two components, namely: taking into account the quality approach and metrology and optimizing the networking of economic actors, higher education establishments, and agronomic and forestry research institutes from the Mediterranean region. Within the frame of the QESAMED project the CAC benefits from several assessments (initial, intermediate, and final), as well as specific support in implementing the quality approach, including trainee tutoring and training.
The methodology adopted to implement this approach includes an initial assessment carried out as part of the QESAMED project, identification of the range of laboratory activities and testing standards to be accredited, consideration of the metrological requirements of these standards, and staff training in ISO/IEC 17025 and metrology. This methodology is detailed below.
2.1 Initial assessment
This first step made it possible to:
Carry out a diagnosis of the existing situation which focused on the internal organization, activities, documentation, and improvement dynamics of the center;
Establish an action plan integrating several actions linked to the various processes, such as establishing the quality commitment and policy, drafting the quality manual, establishing procedures, staff training, etc.
2.2 Identification of the range of laboratory activities to be accredited
A survey was carried out with the Menara Cluster, a group of companies in the agri-food and cosmetics sectors in the Marrakesh region, in order to determine their main needs in terms of physicochemical, microbiological, and organoleptic testing services. A detailed analysis of the survey results showed that 56% of the testing requested was in the field of microbiology. The CAC microbiology laboratory was thus chosen to initiate the quality approach.
2.3 Identification standard testing to be accredited
A more detailed study of needs revealed a number of common testing requested by the majority of future customers. The complete list of the microbiological methods and the target organisms is as follows:
2.4 Equipment: Metrological requirements and management
The quality of measurement results depends on the 5M (material, machine, manpower, method, and mother nature). In this part, we will focus on the “machine” component, which corresponds to the laboratory equipment.
The reliability of the results obtained in a laboratory can only be achieved if measurement and testing equipment are under control. Setting up a metrological function is one way of ensuring this control. According to ISO 10012 [21], the metrological function is a “function with administrative and technical responsibility for defining and implementing the measurement management system”. The metrological function makes it possible to: (i) ensure that measurement and testing equipment meets the requirements corresponding to its intended use, namely by performing operations of calibration, verification, adjustment, repair, etc. (ii) carry out continuous monitoring of measurement processes.
ISO/IEC 17025 [5] has a number of requirements for equipment. The laboratory is obliged to check that the equipment complies with specific requirements (e.g., testing standards) before being put into service or put back into service. It is also necessary to check the need for intermediate equipment checks. Moreover, to ensure the validity of the results, a functional check of the equipment can be carried out, among other things. The equipment used to carry out measurements and testing must be capable of achieving the measurement accuracy and/or uncertainty necessary to deliver a valid result. Therefore, critical equipment must be calibrated according to a calibration program with identification of the calibration status or validity period. These requirements specifically underline the fact that any deviation in equipment affecting the quality of results must be monitored. A small deviation is not necessarily insignificant when it comes to estimating uncertainties [22]. Another requirement of ISO/IEC 17025 [5] highlights the importance of equipment identification, so the laboratory must adopt a unique identification system for its measurement and testing equipment. In practice, one of the most important records to be kept when managing equipment (which may have an influence on laboratory activities) is the life sheet, which may contain a description of all operations (calibration, maintenance, etc.) carried out on the equipment from commissioning to decommissioning because of a malfunction.
At the CAC level, the deployment of the metrology function was initiated by a study of the various testing standards to be accredited. The metrological quantities to be controlled and the associated critical equipment were listed. Critical measurement and testing equipment, with a direct impact on the testing results to be carried out in the laboratory, were identified. This includes the following equipment: incubator, autoclave, water bath, refrigerator, pH meter, balance, and pipette. For each critical equipment, testing standards provide a number of metrological requirements to be met, for example:
Incubator: a programmable value (e.g., 30 °C) plus or minus a tolerance (e.g., ± 1);
pH meter: a resolution (e.g., ± 0.01 pH unit at 25 °C), and a tolerance (e.g., ± 0.1);
Balance: type of balance (analytical balance), resolution (e.g., 1%: to weigh 10 g, the balance should be capable of being read to 0.1 g), and tolerance (± 2 mg);
Water bath: a programmable value (e.g., 47 °C) plus or minus a tolerance (e.g., ± 2 °C);
Pipette: a capacity with a graduation (e.g., 1 ml graduated at 0.1), and the class (e.g., class A) which, for pipettes, represents the tolerance.
Further to these examples, we can see that among the metrological requirements highlighted by testing standards are resolution and tolerance related to quantity (it is tolerance related to the accuracy). This tolerance will subsequently be used to: (i) confirm the metrological consistency of existing laboratory equipment by calculating a capability coefficient; or (ii) determine the maximum permissible error (MPE) of equipment to be purchased.
The metrological requirements of testing standards in relation to the target values to be measured or programmed will make it possible to determine the range of a nominal indication interval of each equipment. This methodology was followed to determine the metrological requirements in relation to all the critical equipment identified. Once the metrological requirements have been determined, the next step is to manage the measurement and testing equipment by establishing a procedure allowing this to be done, including methods for identifying equipment and carrying out calibration and verification operations.
2.5 Staff training
In the field of scientific research, implementing a metrology function is of paramount importance since it provides information about the equipment and therefore facilitates the identification of needs, taking into account the capability of the measurement process. Metrology management increases the confidence placed in results, which can be quantified by measurement uncertainty. For this reason, raising staff awareness of metrology is an essential step in highlighting its benefits and the steps involved in its implementation. This is perfectly in line with the objectives of the QESAMED project, which aims to develop tailored training courses in response to the identified needs for metrology and quality.
It is interesting to note the difficulties associated with training, quality, and metrology on a national scale, namely the lack of academic training and highly specialized teachers in these fields in Morocco. For example, metrology training could take place at universities. This training must have a fundamental aspect, both theoretical and practical. Being an experimental science, metrology needs to offer practical work to train students properly [23].
When implementing a quality management system (QMS) in a research laboratory, and more specifically the metrology, a number of challenges can arise. As pointed out by Molinéro-Demilly et al. [24], one of the challenges encountered is the multidisciplinary nature of the scientific community and the need to get people to work effectively together with different backgrounds and habits, but also with different levels of knowledge and skills related to metrology. It is therefore necessary to create both a common interest and a synergy around metrology and quality within the laboratory. To meet this challenge, it was decided to build a QMS common to all CAC activities. To coordinate the implementation of this system, a quality manager has been recruited.
At CAC level, in order to develop the skills of permanent and non-permanent staff in quality management and metrology, an annual training plan has been established. Three training courses have been organized:
Training in ISO/IEC 17025 [8] standard;
Laboratory metrology training according to ISO/IEC 17025 [8] requirements;
Training in self-evaluation / internal audit according to ISO 19011 [25].
Also, as part of the QESAMED project, two Master's-level internship projects/research projects have been defined, one of which concerns the implementation of the metrology function at the CAC.
On an African scale, one of the ways in which metrology can be developed is through the organization of regional and international events in Africa, and provision of more and more training courses. This can be ensured by the African Metrology Committee (CAFMET) [26] (partner of the QESAMED project), aiming at raising awareness among African organizations (public or private) on the importance of metrology for a country's development by promoting and disseminating a culture of metrology in testing and calibration. This objective can also be achieved by contributing to the sharing of knowledge between technical experts in African countries, and by promoting communication between companies, administrations, higher education institutes, and laboratories dedicated to metrology research and development. CAFMET can also support the implementation of metrological processes, and offer tutorials given by experienced international experts [23].
3 Results
3.1 Equipment metrological requirements
After analysis of the testing standards, the critical quantities are: temperature (°C), mass (g), volume (ml), and pH (pH unit). For each critical measuring and testing equipment, the metrological requirements to be met were determined by choosing the most restrictive metrological requirement of all the testing standards analyzed in the context of this study and which will be used by the laboratory. The results of the analysis are summarized in Table 1.
The next step was to make an inventory of the existing equipment at the laboratory concerned, with the aim of checking the range of a nominal indication interval and confirming the metrological consistency of the equipment. For any equipment required to carry out the testing, and that is non-conforming or non-existent, the technical specifications need to be determined so that it can be purchased.
3.1.1 Identification of the range of a nominal indication interval
By definition, the range of a nominal indication interval is the “absolute value of the difference between the extreme quantity values of a nominal indication interval” [29]. These two extreme values can be the minimum and maximum values that an equipment can measure (e.g., balance, thermometers, etc.) or the extreme values that can be programmed when using an equipment (e.g., autoclave, incubator, etc.). In other words, it is the set of measurand values for which the error of a measuring instrument is assumed to be less than the MPE.
At the CAC level, this step consisted of identifying the range of a nominal indication interval of each equipment required to perform the measurements and testing based on the requirements of the testing standards analyzed. For a balance, for example, the maximum extreme value is referred to as the “maximum capacity”.
3.1.2 Confirmation of the metrological consistency of existing equipment in the microbiology laboratory
Confirmation of the metrological consistency of existing measurement equipment in the microbiology laboratory in relation to the defined metrological requirements was carried out by calculating the capability (C) of each equipment. Capability is measured as the ratio between the required performance and the actual performance of an equipment [30]. It was calculated using equation (1), with a confidence level of the expanded uncertainty of 95% [31].
In this step, we will consider that the only component of uncertainty in the measurement process is the error due to the instrument. Since the equipment has not yet been calibrated (i.e. there is no measurement bias), we'll use the MPE (extreme value of measurement error, with respect to a known reference quantity value, permitted by specifications or regulations for a given measurement, measuring instrument, or measuring system [29]) of the equipment. To evaluate measurement uncertainty, the type B approach using the rectangular distribution is used. The equation (1) for calculating capability then becomes equation (2).
In our case, the minimum capability value to accept is 3. The results of the check of the metrological consistency of measurement and testing equipment are presented in Table 2.
Check of metrological consistency of equipment.
3.1.3 Determining the technical specifications of the equipment to be purchased
The absence or non-compliance of certain equipment has highlighted the need for its purchase. The range of a nominal indication interval, resolution, and MPE are among the most important technical specifications to determine when purchasing measurement or testing equipment. The choice of the MPE is often complex and open to discussion, as it represents a real stake for the cost and reliability of measurement. In the same way as the previous two steps, the range of a nominal indication interval was determined, and the MPE was estimated using the equation (1) with 3 as a capability value. The results of these calculations are shown in Table 3.
Maximum permissible error estimated for equipment to be purchased.
3.2 Management of measurement and testing equipment
Once all the equipment has been inventoried, it needs to be rigorously managed. To this end, a procedure for managing measurement and testing equipment was established. It defines the rules for identifying and monitoring equipment from receipt to decommissioning or reform, including:
Equipment identification by assigning to each equipment a unique identification number composed of an abbreviation referring to the type of equipment and a number assigned according to the order in which equipment of the same type is received over time;
Creation of the equipment file, which includes for each equipment: (i) its life sheet, listing the operations carried out, (ii) its technical data sheet, describing its technical characteristics, (iii) its operating instructions, (iv) its calibration and verification procedures, (v) its internal checks carried out, (vi) its external intervention reports, (vii) its calibration and/or verification certificates, where applicable.
Metrological monitoring of critical equipment is essential to ensure the quality of testing results [33]. ISO/IEC 17025 [8] requires metrological traceability of measurements. To meet the requirements of the standard and ensure the trueness (closeness of agreement between the average of an infinite number of replicate measured quantity values and a reference quantity value [29]) of results, appropriate metrological methods must be chosen to allow linking the result to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty. To identify an efficient strategy (subcontracting the calibration operation to an accredited laboratory or carrying it out in-house), several factors were taken into consideration: the cost of the external metrological service, the competence of the CAC staff, the availability of measurement standards, etc. The frequency initially defined for carrying out metrological operations depends on the criticality of the measurement result, the age of the equipment, the utilization rate, the metrological history, the requirements or recommendations (standards, technical guides, etc.), and finally, the cost [34]. Table 4 summarizes the decisions made for critical equipment.
The frequencies presented in Table 4 may vary over time based on the results of a regular intermediate checks and control charts capable of detecting measurement drifts. These tools can be used to track recalibration equipment needs. Depending on the results obtained, the equipment may not necessarily be recalibrated at fixed time intervals [35].
Methods and frequency of calibration for certain equipment.
4 Discussion
Testing services for external customers are provided to generate an additional budget, and there may be a clear awareness of the need for systematic quality assurance and quality control actions. However, the provision of services is not a high priority in most universities. Staff performance is generally assessed on the basis of published papers and teaching activities, with little or no emphasis on testing services [6]. As a result, the implementation of a quality system often takes a back seat, leading to its postponement. Researchers have stated that a number of characteristics specific to each institution contribute to demotivating, hindering or even preventing the deployment of a quality management system in universities [36]. In the case of the CAC, we have identified a number of challenges related to the nature of the tests to be carried out and others related to the financial management system.
The difficulty of an accreditation process in microbiology is linked, on the one hand, to its technical specificity, combining manual and automated techniques, and, on the other hand, to the lack of information on organizational, technical, and methodological implementation in specialized scientific literature [37]. Among the other obstacles encountered during the realization of this project, some are directly linked to the structural constraints of universities in Morocco. A limited number of permanent staff are permanently assigned to the CAC. One of the transitional solutions put in place is the fact of calling on trainees to carry out testing steps requiring low technical skills, in particular the preparation of culture media and autoclaving operations. Of course, this does not preclude the drafting of a specific procedure describing a solid training, qualification, and authorization process. In the medium term, the recruitment of additional specialized technicians appears to be an absolute necessity. Also, the CAC's financial management is dependent on the strict rules imposed on universities, causing significant delays in the processing of requests for the acquisition of measurement and testing equipment and consumable materials. This has a direct impact on the optimal layout of the microbiology laboratory, in line with accreditation requirements. Attaching the CAC to the City of Innovation, which aims to adopt a status that will allow more flexible financial management, should provide the management flexibility needed to overcome these obstacles.
To facilitate the implementation of a quality management system in accordance with ISO/IEC 17025, the process approach can be used as proposed by some researchers. This approach has a number of advantages: it focuses on the customer; it promotes an overview of all activities and their relationships; it helps to determine responsibilities; it allows processes to be optimized and unnecessary activities to be eliminated; and it generates the elements needed for the evaluation and continuous improvement of the quality management system [38]. To increase the chances of success of such an approach, particular attention must be paid to staff involvement [39,40], and more specifically, trained and qualified laboratory technicians. Without these, the laboratory cannot function effectively [41]. Also, when seeking to accredit a number of tests influenced by random and systematic errors, using different measuring and test equipment, and carried out by staff with different experience and skill levels, risk management and the identification of the impact of changes can be very effective in preventing non-conformities [42].
Finally, there are two possibilities for obtaining accreditation. The competence of testing laboratories can be recognized either by the national organization, the Moroccan Accreditation Service (SEMAC), which is not yet recognized by the International Laboratory Accreditation Cooperation (ILAC), or by an internationally recognized organization such as the French Accreditation Committee (COFRAC), the Tunisian Accreditation Council (TUNAC) or the West African Accreditation System (WAAS).
The definition of the CAC's recognition strategy must take into account the needs of future customers, most of whom wish to export part of their production to Europe. The costs inherent in any request for accreditation, in particular the calling on recognized metrology laboratories to ensure metrological traceability, and the assignment of quality and technical experts from outside the national community for assessments, are thus naturally impacted.
Finally, the implementation of such an approach could, as stated in a study, improve the ability to attract new resources [43].
5 Conclusion
Implementing a quality management system that meets the requirements of ISO/IEC 17025 involves several steps, and requires control of the metrological process in particular. This system allows the laboratory not only to guarantee the reliability and quality of the testing services it provides, but also to identify potential sources of malfunction and improvement. As a result, the impact of this work on the quality organization, internal audits and external evaluation can be significant. Concerning the quality organization, several requirements of the ISO/IEC 17025 standard have been taken into consideration and implemented in relation to the measurement and testing processes. This led to an improvement and formalization of these processes while ensuring effective management based on identification, calibration, verification, and maintenance, thus reducing the risks associated with these processes. Moreover, the definition of a metrological traceability strategy makes it possible to optimize management costs by taking into account national structural constraints. Staff training has led to improved skills, particularly in metrology concepts and standard requirements, which enhances understanding of audit criteria, making audits more effective. Additionally, the documentation and procedures put in place to meet the requirements of the ISO/IEC 17025 standard facilitate the preparation and performance of these audits, during which deviations can be identified, promoting a culture of continuous improvement. The emphasis placed on metrology and traceability of measurements guarantees reliable and reproducible results, essential for the credibility of the microbiological tests carried out. This not only improves the confidence of customers, but also that of the organizations providing recognition during the accreditation process. The latter is necessary for confirming the laboratory's compliance with the standard's requirements, thus strengthening its credibility.
Ultimately, once the various challenges have been overcome, the approach taken will enable the CAC (i) to be equipped with laboratories whose competence to produce technically valid data and results, (ii) to participate in interlaboratory comparisons if they are available and (iii) to undergo an accreditation audit byan independent organization. This will, on the one hand, strengthen CAU's links with the economic sector, particularly the regional food industry, and, on the other hand, reinforce the credibility, on the international scene, of the scientific research activities carried out by teams from the university establishments. Furthermore, international recognition can attract new opportunities for collaboration and funding.
Funding
This research was supported by the European Tempus project n°544564-TEMPUS-1-2013-1-FR-TEMPUS-JPHES.
Conflicts of interest
The authors have no competing interests to declare that are relevant to the content of this paper.
Data availability statement
This article has no associated data.
Author contribution statement
Writing original draft: all authors, Writing-review and editing: Meriem Outaki, Marie-France Thevenon.
References
- Université Cadi AYYAD, https://www.uca.ma/fr [Google Scholar]
- L. Osseni, A. Charki, F. Kebe, G. Calchera, L. Martin, G. Bonnier, Quality management: the challenges of regional governance in West Africa, Int. J. Metrol. Qual. Eng. 6, 404 (2015) [CrossRef] [EDP Sciences] [Google Scholar]
- E. Giesen, Quality management for robust and reliable research, Int. J. Metrol. Qual. Eng. 6, 407 (2015) [CrossRef] [EDP Sciences] [Google Scholar]
- E. Giesen, Ethical and efficient research management: a new challenge for an old problem, Int. J. Metrol. Qual. Eng. 6, 406 (2015) [CrossRef] [EDP Sciences] [Google Scholar]
- A. Rodima, M. Vilbaste, O. Saks, E. Jakobson, E. Koort, V. Pihl, L. Sooväli, L. Jalukse, J. Traks, K. Virro, H. Annuk, K. Aruoja, A. Floren, E. Indermitte, M. Jürgenson, P. Kaleva, K. Kepler, I. Leito, ISO 17025 quality system in a university environment, Accred. Qual. Assur. 10, 369–372 (2005) [CrossRef] [Google Scholar]
- E.A. De Nadai Fernandes, M.A. Bacchi, F.S. Tagliaferro, C.L. Gonzaga, E.J. De França, P.C. Favaro, A. De Angelis Fogaça, Quality system implementation in a Brazilian university laboratory, Accred. Qual. Assur. 10, 594–598 (2006) [CrossRef] [Google Scholar]
- D. Zapata-García, M. Llauradó, G. Rauret, Experience of implementing ISO 17025 for the accreditation of a university testing laboratory, Accred. Qual. Assur. 12, 317–322 (2007) [CrossRef] [Google Scholar]
- NM ISO/IEC 17025 General requirements for the competence of testing and calibration laboratories (IMANOR, 2006) [Google Scholar]
- T. Bourgault, E. Vinner, J.-P. Darchis, C. Arfi, and M. Boutry, Problématique de la mise en place d'une unité de métrologie dans le futur centre de biologie du CHRU de Lille, ITBM-RBM 27, 257–261 (2006) [CrossRef] [Google Scholar]
- C. Naudin, À propos d'une métrologie «raisonnable», Revue Francophone des Laboratoires 2012, 12–13 (2012) [Google Scholar]
- M. Outaki, E. Kerak, Evaluation of implementation of good quality management practices related to the support processes in the laboratories of Moroccan faculties of sciences and techniques, Accred. Qual. Assur. 23, 257–267 (2018) [CrossRef] [Google Scholar]
- ISO 4833-1 Microbiology of the food chain — Horizontal method for the enumeration of microorganisms — Part 1: Colony count at 30 degrees C by the pour plate technique (ISO, 2013) [Google Scholar]
- ISO 4833-2 Microbiology of the food chain — Horizontal method for the enumeration of microorganisms — Part 2: Colony count at 30 degrees C by the surface plating technique (ISO, 2013) [Google Scholar]
- ISO 4832 Microbiology of food and animal feeding stuffs — Horizontal method for the enumeration of coliforms — Colony-count technique (ISO, 2006) [Google Scholar]
- ISO 16649-2 Microbiology of food and animal feeding stuffs — Horizontal method for the enumeration of beta-glucuronidase-positive Escherichia coli — Part 2: Colony-count technique at 44 degrees C using 5-bromo-4-chloro-3-indolyl beta-D-glucuronide (ISO, 2001) [Google Scholar]
- ISO 6888-1 Microbiology of food and animal feeding stuffs — Horizontal method for the enumeration of coagulase-positive staphylococci (Staphylococcus aureus and other species) — Part 1: Technique using Baird-Parker agar medium (ISO, 1999) [Google Scholar]
- ISO 6888-3 Microbiology of food and animal feeding stuffs — Horizontal method for the enumeration of coagulase-positive staphylococci (Staphylococcus aureus and other species) — Part 3: Detection and MPN technique for low numbers (ISO, 2003) [Google Scholar]
- ISO 6579 Microbiology of food and animal feeding stuffs — Horizontal method for the detection of Salmonella spp (ISO, 2002) [Google Scholar]
- ISO 21527-1 Microbiology of food and animal feeding stuffs — Horizontal method for the enumeration of yeasts and moulds — Part 1: Colony count technique in products with water activity greater than 0. 95 (ISO, 2008) [Google Scholar]
- ISO 21527-2 Microbiology of food and animal feeding stuffs — Horizontal method for the enumeration of yeasts and moulds — Part 2: Colony count technique in products with water activity less than or equal to 0.95 (ISO, 2008) [Google Scholar]
- ISO 10012 Measurement management systems — Requirements for measurement processes and measuring equipment (ISO, 2003) [Google Scholar]
- A. Charki, F. Pavese, Data comparisons and uncertainty: a roadmap for gaining in competence and improving the reliability of results, Int. J. Metrol. Qual. Eng. 10, 1 (2019) [CrossRef] [EDP Sciences] [Google Scholar]
- The African Committee of Metrology, A. Charki, C. Ferrier, G. Bonnier, M. Berrada, E.A. Hassani, L. Osseni, Metrology in Africa: role of the CAFMET, Accred. Qual. Assur. 14, 227–230 (2009) [CrossRef] [Google Scholar]
- V. Molinéro-Demilly, A. Charki, C. Jeoffrion, B. Lyonnet, S. O'Brien, L. Martin, An overview of Quality Management System implementation in a research laboratory, Int. J. Metrol. Qual. Eng. 9, 2 (2018) [CrossRef] [EDP Sciences] [Google Scholar]
- ISO 19011 Guidelines for auditing management systems (ISO, 2018) [Google Scholar]
- CAFMET, https://www.cafmet.com/ [Google Scholar]
- ISO 7218 Microbiology of food and animal feeding stuffs — General requirements and guidance for microbiological examinations (ISO, 2007) [Google Scholar]
- ISO 6887-1 Microbiology of the food chain — Preparation of test samples, initial suspension and decimal dilutions for microbiological examination — Part 1: General rules for the preparation of the initial suspension and decimal dilutions (ISO, 2017) [Google Scholar]
- ISO/IEC Guide 99 International vocabulary of metrology — Basic and general concepts and associated terms (VIM) (ISO, 2011) [Google Scholar]
- D. Duret, M. Pillet, Qualité en production de l'ISO 9000 à Six Sigma, 3th edn. (Eyrolles-Éd. d'Organisation, 2005) [Google Scholar]
- ISO/IEC Guide 98-4 Uncertainty of measurement — Part 4: Role of measurement uncertainty in conformity assessment (ISO, 2012) [Google Scholar]
- NF EN 45501 Aspects métrologiques des instruments de pesage à fonctionnement non automatique (AFNOR, 1993) [Google Scholar]
- J.-P. Klein, L'accréditation en bactériologie, Revue Francophone des Laboratoires 41, 39–50 (2011) [CrossRef] [Google Scholar]
- J.-M. Reifenberg, E. Riout, A. Leroy, Métrologie dans un laboratoire de biologie médicale: enjeux et difficultés, Revue Francophone des Laboratoires 2014, 69–76 (2014) [CrossRef] [Google Scholar]
- S.E. Gaber, Adjusting external calibration intervals for auxiliary devices in testing laboratories according to intermediate checks results, Water Sci. 35, 82–88 (2021) [CrossRef] [Google Scholar]
- I.H. Grochau, C.A. Ferreira, J.Z. Ferreira, and C.S. Ten Caten, Implementation of a quality management system in university test laboratories: a brief review and new proposals, Accred. Qual. Assur. 15, 681–689 (2010) [CrossRef] [Google Scholar]
- O. Bouix, P. Sanchez, J.-M. Reifenberg, Définition des erreurs maximales tolérées pour les grandeurs physiques ayant un effet significatif sur le résultat du groupage ABO-RH1 en technique microplaque, Transf. Clinique et Biologique 14, 348–351 (2007) [CrossRef] [Google Scholar]
- I.H. Grochau, C.S. Ten Caten, A process approach to ISO/IEC 17025 in the implementation of a quality management system in testing laboratories, Accred. Qual. Assur. 17, 519–527 (2012) [CrossRef] [Google Scholar]
- S. Martínez-Perales, I. Ortiz-Marcos, J.J. Ruiz, A proposal of model for a quality management system in research testing laboratories, Accred. Qual. Assur. 26, 237–248 (2021) [CrossRef] [Google Scholar]
- L.N. De Jesus, R.B. Penteado, F.C. Malheiros, L.A. Medrano Castillo, L.F.M. De Almeida, The conception and initial years of a quality management system based on ISO/IEC 17025: an action research, Accred. Qual. Assur. 28, 147–157 (2023) [CrossRef] [Google Scholar]
- A. Koomson, J. Ntow, P. Dordunu, E. Birikorang, D.T. Ayitey, Laboratory technology: a “neglected” but unique interdisciplinary tool helping to enhance scientific research and development in academic and research institutions in Africa, Unibulletin 10, 99–115 (2021) [CrossRef] [Google Scholar]
- A. Samimi, World of Researches Publication, Risk Management in the Laboratory based on the 17025 Standards, 7, 114–119 (2020) [Google Scholar]
- M. Poli, E. Bianchini, D. Cornolti, S. Pardini, E. Pisano, E. Simi, L. Tedeschi, G. Iervasi, How does it work a quality/performance management system in a biomedical research institution? Lights and shadows, IJQR 15, 871–888 (2021) [Google Scholar]
Cite this article as: Meriem Outaki, Sara Benhaida, Majdouline Belaqziz, Abdelkader Outzourhit, Gilles Calchera, Marie-France Thévenon, Luc Martin, Abdérafi Charki, Hamid Ajana, Quality approach towards ISO/IEC 17025 accreditation at the center of analysis and characterization of Marrakesh Cadi Ayyad University: metrology stakes, Int. J. Metrol. Qual. Eng. 15, 22 (2024)