Open Access
Table 2
Uncertainty budget for the resistor 10 MΩ.
Uncertainty sources |
Probability distribution | Divider | Uncertainty contribution by voltage, Ω | Uncertainty contribution by current, Ω |
---|---|---|---|---|
Repeatability (Type A) | Normal | 1 | 59 | 144 |
Applied signal calibration (B1) | Normal | 2 | 0.6 | 0.3 |
Calibration of Rs (B2) | Normal | 2 | 45 | 45 |
Drift of Rs (B3) | Rectangular |
![]() |
0.3 | 0.3 |
Temperature coefficient of Rs (B4) | Rectangular |
![]() |
12 | 12 |
DMM ratio accuracy (B5) | Normal | 2 | 1 | 1 |
Combined standard uncertainty (Uc ) | ± 75 Ω | ± 151 Ω | ||
Effective degrees of freedom | ∞ | ∞ | ||
Expanded uncertainty (U exp.) at confidence level 95%, (k = 2): | ± 150 Ω = 15 μΩ/Ω | ± 302 Ω = 30.2 μΩ/Ω |
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.