Table 1
Results of the uut standard resistors.
Nominal resistance value | Applied voltage | Applied current | Actual value | Deviation by applied voltage, μΩ/Ω | Deviation by applied current, μΩ/Ω |
---|---|---|---|---|---|
100 mΩ | 300 mV | 1.5 A | 100.00240 mΩ | −302.9 | −232.4 |
1 Ω | 400 mV | 200 mA | 1.0000007 Ω | −28.0 | −14.9 |
10 Ω | 2 V | 100 mA | 9.999998 Ω | −219.4 | −211.9 |
100 Ω | 10 V | 50 mA | 99.999997 Ω | −35.2 | −34.3 |
1 kΩ | 30 V | 15 mA | 1.0000007 kΩ | 7.7 | 10.0 |
10 kΩ | 30 V | 1.5 mA | 10.0000098 kΩ | 0.8 | 1.1 |
100 kΩ | 30 V | 150 μA | 99.999920 kΩ | 8.2 | 10.5 |
1 MΩ | 30 V | 15 μA | 1.0000008 MΩ | −23.1 | −35.0 |
10 MΩ | 30 V | 1.5 μA | 10.000020 MΩ | 110.5 | 116.5 |