Open Access

Table 2

Uncertainty budget for the measurements.

Source of uncertainty Type A value (%) Type B value (%)
1. Statistical uncertainty
     Repeatability of measurements 0.06
     Reproducibility of measurements 0.35
2. Uncertainty of dosimeter reading that include: calibration factor, resolution of reading, stability and ion recombination 0.17 0.52
3. Uncertainty due to position of ion chamber (uniformity and distance) 0.2 0.2
4. Uncertainty due to the measurements of environmental conditions, KTP (temperature and pressure) 0.35 0.18
5. Uncertainty due to beam scattering 0.35 1

Combined relative standard uncertainty = 1.34%.