Research Article
Development of 137Cs irradiation facility for metrological application
Department of Ionizing Radiation Metrology, National Institute for Standards (NIS),
El-Haram,
Giza, Egypt
* Corresponding author: aelsersy@yahoo.com
Received:
10
November
2016
Accepted:
2
May
2017
In this work, 137Cs irradiation facility at the National Institute of Standards (NIS) was developed to enhance the calibration processes. Different thickness lead sheets were used for beam attenuation, to enlarge the 137Cs dose range. Ambient dose equivalent rate, H*(10) was measured using two ionization chambers at a different source-detector distances (SDDs) and a lead sheet thickness. The deviation from the inverse square law for the dose–SDD relationship was obtained. The beam flatness of unattenuated and the lead attenuated beam was measured. The attenuation coefficient for the used lead sheets was obtained and correlated to the broad beam geometry. The room scattering was studied in detail. The uncertainty of the measured doses was calculated. It was found that the inverse square law is verified well at SDD ≥ 2 m. The measured attenuation coefficient for lead attenuators is affected by the broad beam geometry. The contribution of the scattering component of the dose at a certain SDD is the major source of uncertainty since it extracted a value of 1.05%, while the combined uncertainty from all other factors affecting the dose measurements is 1.34%.
Key words: ionizing radiation / radiation protection / Cs-137 / scattering / irradiation facility
© M.A. Hassan et al., published by EDP Sciences, 2017
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.