Thermoelectric properties of currently available Au/Pt thermocouples related to the valid reference function
1 Physikalisch-Technische Bundesanstalt
(PTB), Abbestraße 2-12, 10587
Berlin,
Germany
2 Turkiye Bilimsel ve Teknolojik
Arastirma Kurumu (TUBITAK), Turkey
3 Centro Español de Metrología (CEM),
C/del Alfar 2, 28760
Tres Cantos, Madrid,
Spain
4 Biroul Roman de Metrologie Legala
(BRML-INM), Romania
5 National Physical Laboratory
(NPL), Hampton Road,
Teddington, Middlesex, UK
6 Institute of Metrology of Bosnia and
Herzegovina (IMBIH), Augusta Brauna
2, 71000
Sarajevo, Bosnia and
Herzegovina
7 Główny Urza¸d Miar, ul. Elektoralna
2, 00-139
Warszawa,
Poland
8 Directorate of Measures and Precious
Metals, Mike Alasa
14, 11000
Belgrade, Republic of
Serbia
9 Cesky Metrologicky Institut Brno
(CMI), Czech Republic
⋆ Correspondence:
frank.edler@ptb.de
Received:
14
March
2015
Accepted:
9
June
2015
Au/Pt thermocouples are considered to be an alternative to High Temperature Standard Platinum Resistance Thermometers (HTSPRTs) for realizing temperatures according to the International Temperature Scale of 1990 (ITS-90) in the temperature range between aluminium (660.323 °C) and silver (961.78 °C). The original aim of this work was to develop and to validate a new reference function for Au/Pt thermocouples which reflects the properties of presently commercially available Au and Pt wires. The thermoelectric properties of 16 Au/Pt thermocouples constructed at different National Metrological Institutes by using wires from different suppliers and 4 commercially available Au/Pt thermocouples were investigated. Most of them exhibit significant deviations from the current reference function of Au/Pt thermocouples caused by the poor performance of the Au-wires available. Thermoelectric homogeneity was investigated by measuring immersion profiles during freezes at the freezing point of silver and in liquid baths. The thermoelectric inhomogeneities were found to be one order of magnitude larger than those of Au/Pt thermocouples of the Standard Reference Material® (SRM®) 1749. The improvement of the annealing procedure of the gold wires is a key process to achieve thermoelectric homogeneities in the order of only about (2–3) mK, sufficient to replace the impracticable HTSPRTs as interpolation instruments of the ITS-90. Comparison measurements of some of the Au/Pt thermocouples against a HTSPRT and an absolutely calibrated radiation thermometer were performed and exhibit agreements within the expanded measurement uncertainties. It has been found that the current reference function of Au/Pt thermocouples reflects adequately the thermoelectric properties of currently available Au/Pt thermocouples.
Key words: Au/Pt thermocouple / reference function / temperature scale