Table 7
The reference values and standard uncertainties at each calibration point calculated using the arithmetic mean of data from the other two labs.
| The serial number of the measuring point i in measurement comparison | 1 | 2 | 3 | 4 | 5 | 6 | 7 | |||
|---|---|---|---|---|---|---|---|---|---|---|
| Correction reference value for lab A humidity comparison YrAi /%RH | –0.4 | –0.5 | 0.0 | 0.6 | 1.3 | 2.3 | 3.0 | |||
| Standard uncertainty of reference value for laboratory A humidity comparison urAi /%RH | 0.7 | |||||||||
| Correction reference value for lab B humidity comparison YrBi /%RH | –0.8 | –0.8 | –0.6 | –0.2 | 0.5 | 1.0 | 1.2 | |||
| Standard uncertainty of reference value for laboratory B humidity comparison urBi /%RH | 0.7 | |||||||||
| Correction reference value for lab C humidity comparison YrCi /%RH | –0.9 | –0.8 | –0.2 | 0.4 | 0.8 | 1.4 | 1.8 | |||
| Standard uncertainty of reference value for laboratory C humidity comparison urCi /%RH | 0.8 |