Table 4

Judgment of the abnormal value of the humidity correction value in each laboratory.

The serial number of the measuring point i in measurement comparison 1 2 3 4 5 6 7
Arithmetic mean value of humidity correction Y¯i/%RH –0.7 –0.7 –0.3 0.3 0.9 1.5 2.0
Experimental standard deviation of correction value by three laboratories s /%RH 0.46 0.29 0.55 0.76 0.81 1.39 1.87
eAi /%RH 0.5 0.3 0.5 0.7 0.9 1.5 2.0
eBi/%RH 0.1 0.2 0.6 0.8 0.7 1.2 1.7
eCi /%RH 0.4 0.2 0.0 0.2 0.1 0.4 0.3
(eji/s)max 1.09 1.04 1.09 1.05 1.11 1.08 1.07
conclusion no outliers