Table 3
Humidity correction value obtained by each laboratory.
| The serial number of the measuring point i in measurement comparison | 1 | 2 | 3 | 4 | 5 | 6 | 7 | |||
|---|---|---|---|---|---|---|---|---|---|---|
| Laboratory A correction value YAi /%RH | –1.2 | –1.0 | –0.8 | –0.4 | 0.0 | 0.0 | 0.0 | |||
| Extended uncertainty of YAi UAi/%RH | 2.3 | |||||||||
| Laboratory B correction value YBi /%RH | –0.6 | –0.5 | 0.3 | 1.1 | 1.6 | 2.7 | 3.7 | |||
| Extended uncertainty of YBi UBi/%RH | 2.1 | |||||||||
| Laboratory C correction value YCi /%RH | –0.3 | –0.5 | –0.3 | 0.1 | 1.0 | 1.9 | 2.3 | |||
| Extended uncertainty of YCi UCi/%RH | 1.9 |