Open Access

Table 22

Uncertainty budget in consumption, Test 1, Lab B.

Type Source of uncertainty Value (U) Unit Distribution k u (xi) Veff uc (xi)
B Barometer uncertainty − Pa 0.06900 kPa Normal 2 0.0345 0.03456
B Lower division of the barometer − Pa 0.00500 kPa Triangular 0.0020
B Pressure gauge uncertainty − P 0.00007 kPa Normal 2 0.0000 0.00204
B Lower division of the pressure gauge − P 0.00500 kPa Triangular 0.0020
B Flow meter uncertainty − 0.00038 m3/h Normal 2 0.0002 0.00011
B Lower division of the flow meter − 0.00005 m3/h Triangular 0.0000
B Uncertainty of the gas thermometer − Tg 0.33000 °C Normal 2 0.1650 0.19401
B Lower division of the gas thermometer − Tg 0.25000 °C Triangular 0.1021
A Repeatability 0.00163 kW Normal 0.0006 6 0.00062