1 Introduction

With the development of science and technics, biosystem is one of the main references to create new technologies. Under this background, bionics has been promoted in developing novel sensor designs, investigating new sensor materials and studying sensing principles. Based on the bionics principle, Xue et al. firstly presented the cilium-shaped MEMS vector hydrophone (CVH) to mimick the lateral organs of fish, and its sensitivity reached up to −197.7 dB at 1 kHz (0 dB re 1 V/µPa) [1]. Bionic cilia cylinder is used to pick up sound signals, which is a significant sensing part for CVH. Inspired by this bionic principle, Hanasoge et al. proposed a magnetic artificial cilia fabricating method, which is applied to pump and mix microfluidic, and other fluid handing processes [2]. Kottapalli et al. reported a SU-8 3D hair cell structure (Fig. 1a) [3]; Krijnen et al. reported the array of spiral-suspended sensory hairs (Fig. 1b) [4]; Zhang et al. reported the design and analysis of a multiple sensor units vector hydrophone (Fig. 1c) [5].

Based on all these above, biomimetic sensors are gradually developing towards miniaturization and arrays. Therefore, North University of China designed and manufactured FUVH based on the research of MUVH. The FUVH design process takes fish lateral line neuromasts as the reference. By simulating this biologic structure and its function, the design purpose of FUVH is to improve the performance of MEMS vector hydrophone for detecting remote ships noise [6]. However, FUVH is still used as a simple combination of four independent cilium MEMS vector hydrophones (CVHs) in current application. Neither the design of sensing structure nor the application methods are innovative. Thus, the advantage of multiple sensing units cannot be fully utilized. Meanwhile, still using previous cilium-shaped sensing structure is unable to achieve the current application purpose of FUVH. From previous research, the other main restriction is that the improvement of sensitivity is counteracted by the expanding bandwidth [7]. Therefore, methods to improve the sensitivity of sensing structures and to ensure the appropriate bandwidth have been presented as follows: Bora et al. reported the HA-MA hydrogel encapsulated PZT sensor that was covered by the dome shaped cupula (Fig. 2a) [8]; Liu et al. presented a “lollipop-shaped” MEMS vector hydrophone, whose cilium structure is integrated with a low density sphere (Fig. 2b2) [9]; Xu et al. presented a “cup-shaped” MEMS vector hydrophone, whose main sensing structure is replaced by an artificial “cup” (Fig. 2b3) [7].

Both the design of lollipop-shape and cup-shape are used to mimick the cupula of the lateral line neuromast, which has been proven to enhance the sensitivity by expanding sensing areas. However, for the size restriction of FUVH sensor chip, both the design methods above are no longer fit for FUVH. Based on this principle, the anulus-shaped ciliary structure for FUVH is proposed in this paper. This novel structure integrates a low-density rigid ring with previous cilium cylinders, which not only makes four single sensor units of FUVH become an entirety, but also ensures the advantages of sensitivity and working bandwidth. Features of different hydrophones are shown in Table 1.

thumbnail Fig. 1

Biomimetic sensors based on a bionic cilium.

thumbnail Fig. 2

Biomimetic sensors based on the expanding sensing areas.

Table 1

Features of different hydrophones.

2 Bio-inspiration

From the study of lateral line system, the signal receptors are neuromasts that are mainly used to perceive the flow of water and the movement of the prey target. Taking superficial neuromasts (SNs) as example, SNs are covered by a slender cupula and have bundles of cilia inside. When lateral line system senses signals, the cupula tilts, then the ciliary bundles transmit stimulus signals to sensory cells, and then to nerve fibers [10], as shown in Figure 3a. The innervation structure of SNs is that a nerve innervates a group of adjacent SNs after branching. As shown in Figure 3b, this structure and SN groups constitute the fish lateral line system together. The SEM image of the blind cave fish lateral line neuromast is shown in Figure 3c.

Based on the SNs bionics principle and hydroacoustic propagation theory, the sensing structure of AFUVH is composed of three parts: a solid ring, four cilia cylinders and cross-beams of four sensor units. All three parts are rigidly intergraded to ensure that the force moment is transmitted without any rotation angle. The rigid ring is a simple but effective simulation of the cupula macroscopically. The whole structure accomplishes the simulation of the sensing process, from the movement of cupula to the signal reception of sensory cells. When annulus-shaped ciliary structure receives underwater acoustic signals, the rigid ring firstly drives four cilia cylinders to move. Their movements transfer the force moment to cross-beams, and then induce the deformation of four cross-beams in each unit. Resistances (rate) of the piezoresistors on the cross-beams change with the deformation. Finally, through Wheatstone Bridge, the acoustic signals are converted to the voltage outputs, which is the complete conversion process of acoustic signals to electrical signals.

A group of multiple SNs respond to stimulus signals by forming sensory functional groups through specific innervation structures, which means the stimulus signals received by each sensory cell in a single SNM can be integrated and transmitted through nerve branches. Based on this bionic principle, annulus-shaped ciliary structure realizes the consistency of signal phases received by four units, which makes FUVH a linear additive array. Thus, the received signals can be directly added to facilitate algorithm processing (Fig. 4).

thumbnail Fig. 3

(a) Four hair cells occur at the center of a schematic depiction of a single SN, modified after Ref. [11]; (b) Schematic diagram of fish lateral line system; (c) The SEM image of the blind cave fish lateral line neuromast: a neuromast consists of five adjacent sensory cells (yellow) and a lateral nerve (green).

thumbnail Fig. 4

(a) Cilium MEMS vector hydrophone (CVH); (b) Four-unit MEMS vector hydrophone (FUVH); (c) Anulus-shaped ciliary structure MEMS vector hydrophone (AFUVH).

3 Mechanical analysis

3.1 Sensing materials selection

AFUVH is a resonant MEMS hydrophone, which mainly relies on the co-vibration of the sensing structure and the acoustic medium to receive acoustic signals. From vibrational science and hydroacoustics, to meet the conditions of picking up vibration [11,12], the average density of the annulus-shaped ciliary structure must be almost equal to that of water medium. Based on this principle, PE and PDMS firstly raise our attention, whose relevant properties are shown in Table 2. The density of polyethylene (PE) and polydimethylsiloxane (PDMS) are both closed to that of water. Meanwhile, the appropriate toughness and rigidity of these two materials enable the external stress be transmitted easily. Therefore, PDMS is chosen to fabricate the rigid ring to form the cupula and PE is chosen to fabricate four cilia cylinders in this work.

Table 2

Material properties.

3.2 Properties of sensing structure

Piezoresistive effect of monocrystal silicon is the sensitive mechanism of AFUVH, by which mechanical quantity is transferred into electrical signals. With the stress increase on the cantilever beam, the resistance (rate) of piezoresistor changes more sharply along with the deformation of the piezoresistance. Then, the output voltage of Wheatstone bridge becomes much higher. That is to say the stress on the cantilever beam reflects the sensitivity of AFUVH.

According to previous research [13], the material mechanics principle and theoretical mechanics analysis, stress σ (x) at any point of a single cantilever beam is expressed as:(1)

Wherein P is the external sound pressure and (β is the coefficient of torsion of cantilever beam section, and μ is the Poisson's ratio of the cantilever beam). Different meanings of the parameters in formula (1) are shown in Figure 5. The parameters of the cross-beam in a single unit of FUVH are the same with previous work, which are shown in Table 3 [14]. Formula (1) indicates that the stress on the cross-beam rises with the radius and height of cilia cylinders, and the height of ring. That means the longer or thicker the cilia cylinders are and the higher the ring is, the higher sensitive of AFUVH is. It is also worth noting that, because of the ring, the barycentre of the whole sensing structure is projected on the center of the sensor chip, which is different from CVH whose barycentre of sensing structure is projected on the center mass block. Thus, there's an initial stress on the cantilever beam, which reduces the stress variation range of the cantilever beam in the inner part of ring while increasing that in the outer part of ring.

Besides, the bandwidth is also an important feature of AFUVH performance, which is mainly decided by the natural frequency of the sensing structure. From previous study, with the increase of natural frequency, the bandwidth of AFUVH becomes wider. In addition, considering the application environment is water, the bandwidth of AFUVH is seriously affected by the fluid-solid coupling (FSI) of the sensing structure and water [15,16]. Thus, the static stress analysis and FSI analysis for AFUVH should be carried out further.

thumbnail Fig. 5

Geometric parameter schematic of AFUVH.

Table 3

Parameters of cross beam.

4 Simulation optimization of AFUVH

In this paper, by using ANSYS 17.0, the static analysis is carried out to indicate the maximum stress on X and Y direction cantilever beams, and FSI analysis is performed to indicate the natural frequency of AFUVH. Design variables are H (the height of cilia cylinders), D (the diameter of cilia cylinders), h (the height of ring) and L (the thickness of ring). Considering the restrictions of manufacturing process and sensor chip parameters, the design and optimization of annulus-shaped ciliary structure are performed by using the following parameter ranges: (1) 2 mm ≤ H ≤ 5 mm; (2) 0.2 mm ≤ D ≤ 0.5 mm; (3) 0.5 mm ≤ h ≤ 3 mm; (4) 0.2 mm ≤ L ≤ 1 mm; (5) FSI frequency f > 1000 Hz. Figures 6 and 7 show the simulation results, and Figure 8 shows the impact of the variable parameters.

As shown in Figure 6, the stress on the cantilever beam rises with the increase of H and h. In Figure 7, except for diameter D, all other parameters have the negative relationship with the sensing structure natural frequency. Figure 8 shows the influence of different parameterson natural frequency and maximum stress, which indicates that the stress on the cantilever beam and the sensing structure natural frequency are a pair of contradictory features. That is to say the sensitivity and the bandwidth of AFUVH are also a pair of contradictory features.

From Figure 8, the parameter D has the great positive influence on natural frequency and the little negative influence on maximum stress on the cantilever beams, while the parameter L are totally on the contrary. Therefore, changing two specific parameters D and L is sufficient to solve the problem above. Then, selecting the appropriate combination of different parameter sizes is of great importance. Two design purposes of the simulation optimization progress are shown as follows: (1) AFUVH has the better sensitivity than FUVH; (2) AFUVH retains its bandwidth more than 1000 Hz as much as possible.

Before determining the specific parameter sizes, establishing a preliminary model of AFUVH and carrying out a static analysis of this model, the stress nephogram is shown in Figure 9a and the stress curves of this preliminary model are shown in Figure 9b. Especially in Figure 9b, it can be seen from the stress curves of four beams in the X direction that the maximum stresses on cantilever beams (1) and (4) are obviously greater than those on cantilever beams (2) and (3).That is to say the stress variation range of the cantilever beams in the inner part of ring is wider than that in the outer part of ring, which is consistent with the mechanical analysis in PART 3.2. Because of the great consistency of sensor chip produced by MEMS process, the resistances of piezoresistors on each cantilever beam are almost same, which means any two cantilever beams in the same direction can be combined to form a Wheatstone bridge. Thus, this problem can be solved by adjusting the combination of cantilever beams. Figure 10 shows the overall FUVH sensor chip. Finally, AFUVH can be combined to four rearranged sensor units (two sensor units with high sensitivity as shown in Fig. 10a, and two with low sensitivity as shown in Fig. 10b), which is different from cilium-shaped FUVH.

Comparing this new sensing structure with cilium-shaped FUVH, the cilium radius of both structures should be invariant. Meanwhile, to ensure the improvement of AFUVH performance, the maximum stress on the cantilever beam in the inner part of ring should be not less than that of FUVH. On these premises, the parameter sizes of AFUVH basic model are determined in this paper: H = 3 mm, D = 0.4 mm, h = 1.75 mm and L = 0.72 mm. In this model, the maximum stress on the cantilever beam in the inner part of the ring is closely equal to that of FUVH. Its design purpose is to demonstrate that AFUVH has better performance than FUVH in receiving acoustic signals.

However, this basic model mainly takes the former design purpose into consideration, which is not comprehensive. Therefore, based on this model, an optimized model of AFUVH is designed by using the multi-parameter optimization function of ANSYS 17.0 workbench and taking the influence factors and influence trends of each parameter as references. The parameter sizes of AFUVH optimized model are finally determined: H = 3 mm, D = 0.4 mm, h = 2 mm and L = 0.2 mm. This optimized model is used to demonstrate what the best effect AFUVH can achieve.

To ensure the reliability of the optimized design method, by using these two model parameters, the static stress analysis and FSI analysis are performed in the following text.

Figure 11 shows the stress contrast curves on the cantilever beam of FUVH, AFUVH basic model, and AFUVH optimized model. Comparing with FUVH, the maximum stress on the cantilever beam of the AFUVH basic model rises up to 2.17 times, which means the sensitivity of a high-sensitivity unit is increased by 6.73 dB. Meanwhile, the maximum stress on the cantilever beam of the AFUVH optimized model rises up to 2.65 times, which means the sensitivity of a high-sensitivity unit of AFUVH can be improved by 8.16 dB. Especially in this structure, the maximum stress on the cantilever beam in the inner part of the ring is 1.39 times that of previous one, which means the sensitivity of a low-sensitivity unit of AFUVH can also be improved by 2.86 dB. In conclusion, the annulus-shaped ciliary structure for FUVH is able to improve the sensitivity of previous cilium-shaped sensing structure for FUVH effectively.

Figure 12 shows the results of FSI analysis. The wet mode response results of AFUVH are both more than 1000 Hz (The first-order wet mode response of AFUVH basic model and AFUVH optimized model reaches up to 1003.93 and 1216.62 Hz, respectively), which achieves the original design purposes. Therefore, the annulus-shaped ciliary structure can improve the performance of FUVH. Then, a prototype of AFUVH are fabricated and tested experimentally.

thumbnail Fig. 6

The relationship between parameters (H, h, R, L) and maximum stress on the cantilever beam.

thumbnail Fig. 7

The relationship between parameters (H, h, R, L) and natural frequency.

thumbnail Fig. 8

The impact of parameters (H, h, R, L) on natural frequency and maximum stress on the cantilever beams.

thumbnail Fig. 9

Stress nephogram and stress curves of AFUVH. (a) Stress nephogram (b) Stress curves of the cantilever beams.

thumbnail Fig. 10

New combination of AFUVH sensor units.

thumbnail Fig. 11

The stress contrast diagram on the cantilever beams.

thumbnail Fig. 12

The analysis of FSI. (a) Fluid-solid coupling interface (b) AFUVH basic model (c) AFUVH optimized model.

5 Experiments for AFUVH

5.1 Fabrication

5.1.1 Fabrication of sensor chips

Four-unit sensor chips are fabricated by MEMS silicon micro-machining technology, which mainly use the bulk micro-machining technology and silicon bond binding technology. The process includes lithography, oxidation, bonding, etching, ion implantation, film growth, evaporation and so on. Four-unit sensor chips are finally formed after deep silicon etching, wet etching and other processes. Figure 13 shows the SEM image of FUVH sensor chip.

thumbnail Fig. 13

SEM image of FUVH sensor chip.

5.1.2 Consistency test of four-unit sensor chip

According to Section 4, Wheatstone bridges formed by the piezoresistors on cantilever beams need to be recombined, which requires great consistency of the piezoresistors manufactured by MEMS processing technology. Thus, it is necessary to test each value of 32 piezoresistors distributed on sixteen cantilever beams.

Taking a unit of FUVH chip as an example, the distribution of piezoresistors on cantilever beams is shown in Figure 14a. Tx1, Tx1, Tx2, Tx2 and Ty1, Ty1, Ty2, Ty2 respectively represent the resistance testing points (“X” means X-direction and “Y” means Y-direction). Figure 14b shows the connection diagram of equivalent resistances, in which the design value of each equivalent resistance is R = 500 Ω. Thus, the equivalent resistance between VCC and Tx1 can be represented as R VccTx1, and the rest equivalent resistances between VCC and other testing points are represented in the same way. Theoretically, the design value of R VccTx1, R VccTx2, R VccTy1, R VccTy2 should be 2R = 1000 Ω and R VccTx1, R VccTx2, R VccTy1, R VccTy2 should be R = 500 Ω. The testing value of equivalent resistances are shown in Table 4.

From Table 4, the testing value of 2R is about 1000 ± 31 Ω and R is about 500 ± 16 Ω, which means the maximum inconsistency (maximum error of R) of piezoresistors on cantilever beams is about 3.2%. In conclusion, the FUVH sensor chip is of good consistency.

thumbnail Fig. 14

Testing method of piezoresistors on the cantilever. (a) Distribution of piezoresistors on the cantilever beams in a unit (b) the connection of equivalent resistances.

Table 4

Testing value of equivalent resistances.

5.1.3 Fabrication of sensing structure

Considering the accuracy of the annulus-shaped ciliary structure fabricating process, an accurate casting model method is adopted. Firstly, infusing the PDMS into a mold, after the solidification of the PDMS, the rigid ring is completed. Secondly, another key fabrication procedure is the integration of the rigid ring, four cilia cylinders and four center blocks of sensor chip. Through the dedicated integration platform, the UV-curable adhesive is adopted to attach four cilia cylinders to the rigid ring. Then, the integration of the anulus-shaped ciliary structure and cross beam is performed in the same way. Thirdly, by taking the way of vacuum vapor deposition, the sensing structure and the four-unit chip are coated with a parylene membrace layer to ensure its overall insulativity [17,18]. Figures 15a–15e shows the fabrication process of the sensing structure and the integrating progress. Figures 16a and 16b shows the actual progresses.

AFUVH is applied to different depth of underwater environment, so it should be encapsulated in a stainless-steel shell from being damaged by the hydraulic pressure. Low-noise preamplifier circuit and the adding circuit are also placed in this shell to pre-process electrical signals from different Wheatstone bridges. Figure 17 shows the material object of AFUVH.

thumbnail Fig. 15

Integration process of the sensing structure.

thumbnail Fig. 16

Integration progress diagram.

thumbnail Fig. 17

Material object of hydrophone.

5.2 Experiments

A specific calibration system is used to test the sensitivity of AFUVH. Figure 18a shows the schematic diagram of calibration system. Figure 18b shows the experimental testing environment. The method of comparative calibration in the standing wave tube is still adopted for the test of sensitivity [19]. The sensitivity can be expressed as:(2)

In formula (2), MX represents the sensitivity of the tested AFUVH; M0 represents the sensitivity of a standard hydrophone (standard hydrophone's sensitivity is −170 dB); V0 and VX respectively represent the output voltage of AFUVH and that of the standard hydrophone; k is the wave number; d and d0 respectively represent the distance from tested AFUVH and standard hydrophone to water surface. According to the sensitivity gain analysis of multiple sensor units vector hydrophone [5], through calculation, frequency response comparison diagram of the three structures are shown in Figure 19.

thumbnail Fig. 18

Experimental test picture. (a) Schematic diagram of calibration system (b) Experimental testing environment.

thumbnail Fig. 19

Sensitivity curve of hydrophone.

6 Discussion

Comparing with cilium-shaped FUVH, the sensitivity of AFUVH basic model can be improved by 3.92 dB at 1000 Hz, in addition, the sensitivity of AFUVH optimized model can be improved by 5.87 dB at 1000 Hz. And there are similar improvements at 20–1000 Hz. These experimental results are consistent with the theory and simulation. Moreover, the working bandwidth of these two model are both above 1000 Hz, which is consistent with the results of FSI simulation and meets the original design requirement. It has been proved that AFUVH has better characteristics than FUVH, but its fabrication process is more complex. The ring structure must be precise and the anulus-shaped ciliary structure must be integrated accurately.

7 Conclusion

In this paper, anulus-shaped ciliary structure for four-unit MEMS vector hydrophone changes the application situation where each unit of four-unit sensor chip is mutually independent. Inspired by bionics, this innovative structure realizes the complete simulation of the lateral line neuromasts of the blind cave fish. By mimicking the biological structure “cupula”, this structure enhances the sensitivity and ensures the appropriate bandwidth at the same time, which improves the performance of four-unit MEMS vector hydrophone. The above judgment is supported by theory, simulation and experiment. The sensitivity and working frequency of AFUVH is measured accurately in the lab. The sensitivity of AFUVH can reach up to 3.92 dB with effective structure and 5.87 dB with optimum dimension structure at 1000 Hz, and satisfies the increments of 6 dB per octave. Meanwhile, the range of AFUVH working frequency is 20–1000 Hz. In conclusion, anulus-shaped ciliary structure is a satisfactory design for four-unit MEMS vector hydrophone, also providing a great reference for further research and application of multi-unit MEMS vector hydrophone.

Acknowledgments

This work was supported by National Natural Science Foundation of China (61525107, 51875535, 61927807), National Key Research and Development Project (2019YFC0119800), The Fund for Shanxi ‘1331 Project’ Key Subject Construction and Innovation Special Zone Project, The Science and Technology Innovation Project of Shanxi Higher Education Institution (2020L0718), The Teaching Reform Project in Higher School in Shanxi (J2020383), Cooperative education project of industry university cooperation (201901157012).

References

  1. C.Y. Xue, S. Chen, W.D. Zhang, B.Z. Zhang, G.J. Zhang, H. Qiao, Design, fabrication, and preliminary characterization of a novel MEMS bionic vector hydrophone, Microelectron. J. 38 , 1021–1026 (2007) [Google Scholar]
  2. S. Hanasoge, M. Ballard, P.J. Hesketh, A. Alexeev, Asymmetric motion of magnetically actuated artificial cilia, Lab on A Chip 17 , 3138–3145 (2018) [Google Scholar]
  3. A.G.P. Kottapalli, M. Bora, E. Kanhere, M. Asadnia, J.M. Miao, M.S. Triantafyllou, Cupula-inspired hyaluronic acid-based hydrogel encapsulation to form biomimetic MEMS flow sensors, Sensors 17 , 14 (2017) [Google Scholar]
  4. G.J.M. Krijnen, M. Dijkstra, J.J. van Baar, S.S. Shankar, W.J. Kuipers, R.J.H. de Boer, D. Altpeter, T.S.J. Lammerink, R. Wiegerink, MEMS based hair flow-sensors as model systems for acoustic perception studies, Nanotechnology 17 , S84–S89 (2006) [PubMed] [Google Scholar]
  5. X.Y. Zhang, Q.D. Xu, G.J. Zhang, N.X. Shen, Z.Z. Shang, Y. Pei, J.W. Ding, L.S. Zhang, R.X. Wang, W.D. Zhang, Design and analysis of a multiple sensor units vector hydrophone, AIP Adv. 8 , 9 (2018) [Google Scholar]
  6. Z. Xiaoyong, L. Laiyuan, Ship radiated noise detector when line spectrum parameters are unknown, Acta Acust. 40 , 511–518 (2015) [Google Scholar]
  7. W. Xu, Y. Liu, G.J. Zhang, R.X. Wang, C.Y. Xue, W.D. Zhang, J. Liu, Development of cup-shaped micro-electromechanical systems-based vector hydrophone, J. Appl. Phys. 120 , 8 (2016) [Google Scholar]
  8. M. Bora, A.G.P. Kottapalli, J.M. Miao, M.S. Triantafyllou, Fish-inspired self-powered microelectromechanical flow sensor with biomimetic hydrogel cupula, APL Mater. 5 , 6 (2017) [Google Scholar]
  9. Y. Liu, R.X. Wang, G.J. Zhang, J. Du, L. Zhao, C.Y. Xue, W.D. Zhang, J. Liu, “Lollipop-shaped” high-sensitivity Microelectromechanical Systems vector hydrophone based on Parylene encapsulation, J. Appl. Phys. 118 , 7 (2015) [Google Scholar]
  10. A. Nagiel, D. Andor-Ardo, A.J. Hudspeth, Specificity of afferent synapses onto plane-polarized hair cells in the posterior lateral line of the zebrafish, J. Neurosci. 28 , 8442–8453 (2008) [PubMed] [Google Scholar]
  11. M.B. Moffett, D.H. Trivett, P.J. Klippel, P.D. Baird, A piezoelectric, flexural-disk, neutrally buoyant, underwater accelerometer, IEEE Trans. Ultrason. Ferroelectr. Fr eq. Control 45 , 1341−1346 (1998) [Google Scholar]
  12. V.A. Shchurov, V.I. Ilyichev, V.P. Kuleshov, M.V. Kuyanova, The interaction of energy flows of underwater ambient noise and a local source, J. Acoust. Soc. Am. 90 , 1002–1004 (1991) [Google Scholar]
  13. B.Z. Zhang, H. Qiao, S. Chen, J. Liu, W.D. Zhang, J.J. Xiong, C.Y. Xue, G.J. Zhang, Modeling and characterization of a micromachined artificial hair cell vector hydrophone, Microsyst. Technol. 14 , 821–828 (2008) [Google Scholar]
  14. M.R. Liu, G.J. Zhang, X.P. Song, Y. Liu, W.D. Zhang, Design of the monolithic integrated array MEMS hydrophone, IEEE Sens. J. 16 , 989–995 (2016) [Google Scholar]
  15. G.J. Zhang, P. Zhao, W.D. Zhang, Resonant frequency of the silicon micro-structure of MEMS vector hydrophone in fluid-structure interaction, AIP Adv. 5 , 8 (2015) [Google Scholar]
  16. C. Vančura, I. Dufour, S.M. Heinrich, F. Josse, A. Hierlemann, Analysis of resonating microcantilevers operating in a viscous liquid environment, Sens. Actuators A Phys. 141 , 43–51 (2008) [Google Scholar]
  17. A. Qualtieri, F. Rizzi, G. Epifani, A. Ernits, M. Kruusmaa, M.D. Vittorio, Parylene-coated bioinspired artificial hair cell for liquid flow sensing, Microelectron. Eng. 98 , 516–519 (2012) [Google Scholar]
  18. N. Chen, C. Tucker, J.M. Engel, Y. Yang, C. Liu, Design and characterization of artificial haircell sensor for flow sensing with ultrahigh velocity and angular sensitivity, J. Microelectromech. Syst. 16 , 999–1014 (2007) [Google Scholar]
  19. W.D. Zhang, L.G. Guan, G.J. Zhang, C.Y. Xue, K.R. Zhang, J.P. Wang, Research of DOA estimation based on single MEMS vector hydrophone, Sensors 9 , 6823–6834 (2009) [Google Scholar]

Cite this article as: Xiaoyong Zhang, Nixin Shen, Qingda Xu, Yu Pei, Yuqi Lian, Weidong Wang, Guojun Zhang, Wendong Zhang, Design and implementation of anulus-shaped ciliary structure for four-unit MEMS vector hydrophone, Int. J. Metrol. Qual. Eng. 12, 4 (2021)

All Tables

Table 1

Features of different hydrophones.

Table 2

Material properties.

Table 3

Parameters of cross beam.

Table 4

Testing value of equivalent resistances.

All Figures

thumbnail Fig. 1

Biomimetic sensors based on a bionic cilium.

In the text
thumbnail Fig. 2

Biomimetic sensors based on the expanding sensing areas.

In the text
thumbnail Fig. 3

(a) Four hair cells occur at the center of a schematic depiction of a single SN, modified after Ref. [11]; (b) Schematic diagram of fish lateral line system; (c) The SEM image of the blind cave fish lateral line neuromast: a neuromast consists of five adjacent sensory cells (yellow) and a lateral nerve (green).

In the text
thumbnail Fig. 4

(a) Cilium MEMS vector hydrophone (CVH); (b) Four-unit MEMS vector hydrophone (FUVH); (c) Anulus-shaped ciliary structure MEMS vector hydrophone (AFUVH).

In the text
thumbnail Fig. 5

Geometric parameter schematic of AFUVH.

In the text
thumbnail Fig. 6

The relationship between parameters (H, h, R, L) and maximum stress on the cantilever beam.

In the text
thumbnail Fig. 7

The relationship between parameters (H, h, R, L) and natural frequency.

In the text
thumbnail Fig. 8

The impact of parameters (H, h, R, L) on natural frequency and maximum stress on the cantilever beams.

In the text
thumbnail Fig. 9

Stress nephogram and stress curves of AFUVH. (a) Stress nephogram (b) Stress curves of the cantilever beams.

In the text
thumbnail Fig. 10

New combination of AFUVH sensor units.

In the text
thumbnail Fig. 11

The stress contrast diagram on the cantilever beams.

In the text
thumbnail Fig. 12

The analysis of FSI. (a) Fluid-solid coupling interface (b) AFUVH basic model (c) AFUVH optimized model.

In the text
thumbnail Fig. 13

SEM image of FUVH sensor chip.

In the text
thumbnail Fig. 14

Testing method of piezoresistors on the cantilever. (a) Distribution of piezoresistors on the cantilever beams in a unit (b) the connection of equivalent resistances.

In the text
thumbnail Fig. 15

Integration process of the sensing structure.

In the text
thumbnail Fig. 16

Integration progress diagram.

In the text
thumbnail Fig. 17

Material object of hydrophone.

In the text
thumbnail Fig. 18

Experimental test picture. (a) Schematic diagram of calibration system (b) Experimental testing environment.

In the text
thumbnail Fig. 19

Sensitivity curve of hydrophone.

In the text

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.