Acceptance criteria established in MSA manual.
Characteristic | Acceptability criteria |
---|---|
Bias and linearity | In general, significantly equal to zero and not exceed the maximum permissible error established by the gage calibration procedure. |
%GRR | <10% tolerance (or process variation): Generally considered to be an acceptable measurement system. |
10–30% tolerance (or process variation): May be acceptable for some applications, but should be approved by the customer. | |
Number of distinct categories (ndc) | Should be greater than or equal to 5. |