Table 1

Uncertainty budget at 0 °C for the two generators.

Source Symbol Distribution Divisor Sensitivity coefficient Value for (1-T) generator (°C) Value for (2-T) generator (°C) Method of estimation
Type A
Repeatability for 1st SPRT 1st SPRT Rep. Normal 1 1 0.009 0.0013 Standard deviation of measurements
Repeatability for 2nd SPRT 2nd SPRT Rep. Normal 1 1 0.0002 Standard deviation of measurements
 
Type B
1st SPRT calibration U(ts sprt cal) Normal 2 1 0.002 0.002 Calibration certificate
SPRT drift U(ts sprt drift) Rectangular √3 1 0 0 Max. difference between two successive calibrations of the standard
1st bath stability U(ts stab) Rectangular √3 1 0.006 0.006 Difference between maximum and minimum value of measurements
2nd SPRT calibration U(ts sprt cal) Normal 2 1 0.002 Calibration certificate
2nd bath stability U(ts stab) Rectangular √3 1 0.006 Difference between maximum and minimum value of measurements
SPRT resolution (indicator unit) U(ts ind res) Rectangular 2√3 1 0.0001 0.0001 Digital indicator = 0.5 of resolution
Bridge calibration U(ts cal) Normal 2 1 0.0005 0.0005 Calibration certificate
Saturator efficiency U(ts sat eff) Normal 1 1 −0.1413 0.0041 The temperature difference between bath temperature that containing the saturator and the air temperature inside the saturator
Saturator homogeneity U(ts sat hom) Normal 1 1 0.01 The temperature difference between saturator inlet and exit
Saturator temperature stability U(ts sat stab) Rectangular √3 1 0.0047 Difference between maximum and minimum value of measurements
Test-chamber efficiency U(ts cham eff) Normal 1 1 0.0057 The temperature difference between bath temperature that containing the chamber and the air temperature inside the saturator
Test-chamber homogeneity U(ts cham hom) Normal 1 1 0.01 The temperature difference between chamber inlet and exit
Test-chamber temperature stability U(ts cham stab) Rectangular √3 1 0.0005 Difference between maximum and minimum value of measurements
Dew-point hygrometer U(tm cal) Normal 2 1 0.06 0.06 Calibration certificate
Resolution of hygrometer U(tm hygro res) Rectangular 2√3 1 0.01 0.01 Digital indicator = 0.5 of resolution
Drift of hygrometer U(tm hygro dri) Rectangular √3 1 0.01 0.01 Max. difference between two successive calibrations of the standard
 
Combined uncertainty         0.073 0.033  
Expanded uncertainty (k = 2)         ±0.145 ±0.066  