Open Access

Table 3

Main specifications of the new fabricated capacitance box.

Item Value/description
Operating power 12 VDC
Operating frequency range <10 kHz
Maximum measuring voltage 75 VAC (peak)
Number of decades/box 3 decades
Number of capacitive elements/decade 4 capacitors instead of 10 in the old decades
Nominal values of the used standard capacitors 100 nF, 200 nF, 400 nF, 800 nF, 1 μF, 2 μF, 4 μF, 8 μF, 10 μF, 20 μF, 40 μF, and 80 μF
Type of the used capacitive components Silvered mica for (100 nF to 1 μF) and metallized polypropylene sulfide for >1 μF
Number of capacitance steps/decade 15 instead of 10 in the old decades
Capacitance steps of each decade 100 nF for the first decade, 1 μF for the second decade, and 10 μF for the third decade
Whole output capacitance range 100 nF to 150 μF
Maximum setting value 166.5 μF
Maximum possible output capacitance steps 4096 instead of 1331 in the old corresponding box
Maximum different output capacitance steps 1666 instead of 1111 in the old corresponding box
Operating temperature and humidity 10 °C to 40 °C; <80% RH
Storage temperature −20 °C to 65 °C
Basic test conditions 1 kHz, 1 V rms, 23 °C
Temperature coefficient 10 to 25 ppm/°C, depending on the capacitance value
Dissipation factor 85–400 ppm depending on the capacitance value
Relative deviation of the output capacitance steps ±5 × 10−5 to ±5 × 10−4
Estimated stability over one year 80–120 ppm/year, depending on the capacitance value
Relative voltage dependence ±1×10−6/V
Measurements repeatability In the order of 1 ppm, depending on the measured capacitance value
Relative uncertainty due to the combination effect <15 ppm, on the average
Box size 20.5 cm × 12.5 cm × 10.5 cm for control unit and 32.5 cm × 27.5 cm × 10.5 cm for the unit of the standards
Mass <4.7 kg

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.

Initial download of the metrics may take a while.